RGS-2000NG TCAS Test Set Operation Manual
1-2-5
September 2021
Subject to Export Control. See Cover Page for details.
Page 183
5.11.17.4.6
Pulse Repetition Frequency (PRF)
5.11.17.4.7
Pulse Width
5.11.17.4.8
PRF Synchronization
Command Syntax:
{:RGS | :RGS2000NG}{:XPDR}{:PRF }SP<numeric>CR
Description:
This command sets the pulse repetition rate (PRF).
Numeric:
1 to 10000 (decimal ASCII)
Default:
100
Example:
:RGS:XPDR:PRF 150\r
Query:
:RGS:XPDR:PRF?\r
Return:
150
Command Syntax:
{:RGS | :RGS2000NG}{:XPDR}{:PWIDTH | :PW }SP<numeric>CR
Description:
This command sets the pulse width.
Numeric:
0 to 5 µs (decimal ASCII)
Default:
0
Example:
:RGS:XPDR:PW 1\r
Query:
:RGS:XPDR:PW?\r
Return:
1.000
NOTE:
This command is valid only for the transponder interrogation mode Pulse.
Command Syntax:
{:RGS | :RGS2000NG}{:XPDR}{:PRFSYNC}SP{ON | OFF}CR
Description:
This command turns on or off the synchronization output.
Default:
Off
Example:
:RGS:XPDR:PRFSYNC ON\r
Query:
:RGS:XPDR:PRFSYNC?\r
Return:
ON
NOTE:
This command is valid only for the following transponder interrogation
modes: P1 to P2, Pulse, DME 12 µs Spacing and DME 30 µs Spacing.