Operation,
Screen Layout and Navigation
1-2-3
RGS-2000NG TCAS Test Set Operation Manual
139187 / Rev. 500
Page 14
Subject to Export Control. See Cover Page for details.
September 2021
3.4.4.2
Measurements - Scope Mode Trigger Softkey Menu
The TCAS Scope Trigger Softkey Menu is accessed by pressing the Trigger Softkey. Trigger settings
can be configured using the Trigger Softkey menu or the fields on the Main Display Area.
SCREEN SEQUE NCE:
TCAS Screen > Measurements Softkey > Setting: Scope Mode > Trigger Softkey
F i g u r e 1 . 2. 3 - 1 4 T C A S Sc o p e Mo de T ri g g e r S o f t ke y s
Screen Components
Description
Source Menu
Selects the Trigger Source.
Antenna Toggle Switch
Selects the Trigger Antenna.
Mode Toggle Switch
Selects the Trigger Mode.
Trigger Softkey
Pressing this softkey clears a triggered event when Trigger Mode is set
to Single.
Level Field
Defines the Trigger Level. The Trigger Level can also be defined on the
Main Display Area using the Trigger Level Slider.
Edge Toggle Switch
Selects the Trigger Edge.