Operation,
Remote Operation
1-2-5
RGS-2000NG TCAS Test Set Operation Manual
139187 / Rev. 500
Page 170
Subject to Export Control. See Cover Page for details.
September 2021
5.11.17.2.4
Mode S Interrogation Message Data
5.11.17.2.5
P1 to P1 Spacing
5.11.17.2.6
Pulse Power Level
Command Syntax:
{:RGS | :RGS2000NG}{:XPDR}{:DBL:}<interrogation
number>{:UF}SP<numeric>CR
Description:
This command sets the data message for the Mode S Interrogation of the
selected double interrogation. <interrogation number> defines the double
interrogation number.
Interrogation:
1 to 2 (decimal ASCII)
Numeric:
Short interrogation 0 to FFFFFFFFFFFFFF (14 hexadecimal ASCII)
Long interrogation 0 to FFFFFFFFFFFFFFFFFFFFFFFFFFFF ( 28
hexadecimal ASCII). The last six characters are the Mode S Address.
Default:
00000000000001
Example:
:RGS:XPDR:DBL:2:UF 5AC4727338FF22\r
Query:
:RGS:XPDR:DBL:2:UF ?\r
Return:
AC4727338FF22
Command Syntax:
{:RGS | :RGS2000NG}{:XPDR}{:DBL}{:P1TOP1}SP<numeric>CR
Description:
This command sets the interrogations spacing from P1 of the first
interrogation to P1 of the second interrogation.
Numeric:
0 to 400 µs (decimal ASCII)
Default:
0
Example:
:RGS:XPDR:DBL:P1TOP1 44\r
Query:
:RGS:XPDR:DBL:P1TOP1?\r
Return:
44
Command Syntax:
{:RGS |:RGS2000NG}{:XPDR}{:DBL:}<interrogation number>{:POW
|:POWER}SP<numeric>CR
Description:
This command sets the antenna power for the top antenna of the selected
double interrogation. <interrogation number> defines the double
interrogation number. The power level range depends on the power mode
selected using the previous command. The bottom antenna power uses this
value plus the antenna power deviation setting
.
Numeric:
-90 to -20 dBm (decimal ASCII)
Default:
-20
Example:
:RGS:XPDR:DBL:1:POW -31\r
Query:
:RGS:XPDR:DBL:1:POW?\r
Returns: