C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-23
set stream time
value
This command enables
value
, where
value
is to attach or disable time stamp
to streaming data string, according to
Table B–5
. The example below
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5.
Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal o3 bkg
This command will auto-calibrate the O
3
background. The example below
shows a successful auto-calibration of the O
3
background. This command is
only available if the photometer option is installed.
Send:
set cal o3 bkg
Receive:
set cal o3 bkg ok
set cal o3 coef
This command will auto-calibrate O
3
coefficients based on O
3
span gas
concentrations. If the mode is incorrect, the instrument responds with
“can’t, wrong settings”. The example below shows a successful auto-
calibration of the O
3
coefficient. This command is only available if the
photometer option is installed.
Send:
set cal o3 coef
Receive:
set cal o3 coef ok
lamp
This command reports the current photometer lamp setting in percent.
The example below shows that the lamp setting is 10.0%. This command is
only available if the photometer option is installed.
Send:
lamp
Receive: lamp
10.0
%
set lamp
value
This command sets the photometer lamp setting to
value
, where
value
is a
floating point number representing a percentage of the photometer lamp
setting. The example below sets the photometer lamp to 10.0%. This
command is only available if the photometer option is installed.
Calibration
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...