Operation
Service Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-85
FLOW CONTROLLER SELECTION:
>ZERO SCALE 10 SLM
GAS 100 SCCM
OPER MODE DIAGS ALARM
The Zero-Air Flow Controller screen is used to select the zero-air
controller. Allowable selections are: 5, 10, 15, and 20 slm. The zero-air
flow controller calibration points are reset to defaults when a new controller
is selected.
Note
Changing the flow controller selection may affect allowable limits for
span concentration and total flow. If is advisable to check these settings.
▲
The mass flow controllers are accurate to only
±
2% upon delivery from the
manufacturer. The instrument includes factory-set calibration curves to
improve this accuracy. To achieve an equivalent accuracy when installing
new controllers, see “
Zero Flow Calibration
” later in this chapter.
▲
●
In the Main Menu, choose Service > Flow Control Selection >
Zero
Scale.
ZERO AIR FLOW CONTROLLER:
CURRENTLY: 5 SLM
SET TO: 20 SLM ?
©ª
CHANGE VALUE
SAVE VALUE
OPER MODE DIAGS ALARM
The Gas Flow Controller screen is used to select the gas controller.
Allowable selections are: 25, 50, 100, 200, and 500 sccm. The gas flow
controller calibration points are reset to defaults when a new controller is
selected.
Zero-Air Flow Controller
Gas Flow Controller
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...