Installation
Remote Interconnections
Thermo Fisher Scientific
Model 146
i
Instruction Manual
2-7
Pins 3, 4, and 23 select a gas. Pins 2, 5, 10, 13, 16, 18, 19, 22, 26, 32, 35,
and 37 are ground.
Note
Gases D, E, and F are available only when the instrument is equipped
with the 6-gas option.
Pins 24, 6, and 25 select a gas span level gas.
Pins 7, 8, 27, and 9 select an ozonator or permeation oven level gas.
Gas Bit 3
Pin 23
Gas Bit 2
Pin 4
Gas Bit 1
Pin 3
Gas Type
0
0
0
Gas Off
0
0
1
Gas A
0
1
0
Gas B
0
1
1
Gas C
1
0
0
Gas D
1
0
1
Gas E
1
1
0
Gas F
Span Bit 3
Pin 25
Span Bit 2
Pin 6
Span Bit 1
Pin 24
Span Level
0
0
0
Span 0
0
0
1
Span 1
0
1
0
Span 2
0
1
1
Span 3
1
0
0
Span 4
1
0
1
Span 5
1
1
0
Manual
Oz/Perm Bit 4
Pin 9
Oz/Perm Bit 3
Pin 27
Oz/Perm 2
Pin 8
Oz/Perm Bit 1
Pin 7
Ozonator or
Permeation
Oven Levell
0
0
0
0
Ozon/Perm Off
0
0
0
1
Ozon Manual
0
0
1
0
Ozon 1
0
0
1
1
Ozon 2
0
1
0
0
Ozon 3
0
1
0
1
Ozon 4
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...