Operation
Service Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-95
●
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Zero
. (Hook up a voltage source of 0 V to the
analog input channel.)
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.12 V ?
CALIBRATE INPUT TO ZERO
OPER MODE DIAGS ALARM
The Analog Input Calibration Full-Scale screen is used to calibrate the full-
scale state of the selected analog input.
●
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Full Scale
. (Hook up a voltage source of 10 V to
the analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 6.24 V
SET TO: 10.0
0
V ?
CALIBRATE INPUT TO ZERO
OPER MODE DIAGS ALARM
The Permeation Oven Settings menu is used for setting up and calibrating
the permeation oven option. The permeation oven settings are visible only
if the permeation oven option installed and when the calibrator is in service
mode. For more information on the service mode, see “
Service Mode
”
earlier in this chapter.
●
In the Main Menu, choose Service >
Perm Oven Settings
.
Note
This adjustment should only be performed by an instrument service
technician.
▲
Analog Input Calibrate
Full-Scale
Permeation Oven
Settings
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...