Operation
Flow Modes
3-20
Model 146
i
Instruction Manual
Thermo Fisher Scientific
than the acceptable range, “TOO LOW” or TOO HIGH” appears on the
screen.
●
In the Main Menu, choose Flow Modes > O3 > GPT/Photometer
Control > Gas A-F > Level 1-5 > Manual >
Zero Air.
MAN GAS A ZERO AIR FLOW:
MINIMUM: 00400 SCCM
MAXIMUM: 10000 SCCM
SET TO: 0100
1
SCCM?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Manual Gas Flow screen sets the manual gas flow value for the
previously selected gas. The screen identifies the minimum and maximum
acceptable gas flow rates and lets the user select a flow rate between these
two settings. If the selected gas flow is lower or higher than the acceptable
range, “TOO LOW” or TOO HIGH” appears on the screen.
●
In the Main Menu, choose Flow Modes > O3 > GPT/Photometer
Control > Gas A-F > Level 1-5 > Manual >
Gas.
MAN GAS A GAS FLOW:
MINIMUM: 004.00 SCCM
MAXIMUM: 100.00 SCCM
SET TO: 050.0
1
SCCM?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Photometer flow mode menu is used to select the photometer
measurement mode. Available choices include: PHOT OFF, PHOT INT,
and PHOT EXT. Ozonator control mode may not be selected from this
screen, and must be selected via the ozonator screen. If the instrument is in
control mode a warning will display “IN OZONATOR CONTROL
MODE”. Once a choice is selected the screen reports that the new mode
has been accepted and activated. The message “SAVING” is displayed for 3
seconds, then removed.
●
In the Main Menu, choose Flow Modes >
Photometer.
Gas
Photometer
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...