Servicing
LCD Module Replacement
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-29
Front panel board
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the three ribbon cables and the two-wire connector from the
front panel board.
3. Pop off the board from the two top mounting studs and remove the
board by lifting it up and off the slotted bottom support.
4. Replace the front panel board by following previous steps in reverse.
Figure 7-11.
Replacing the Front Panel Board and the LCD Module
LCD Module
Replacement
Use the following procedure to replace the LCD module (
Equipment Required:
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Ribbon Cables (3)
Two-Wire Connector
Mounting Studs (2)
Bottom Support
LCD Module Screws (2)
Front Panel Board
LCD Module
Retaining Screw
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...