C-Link Protocol Commands
Measurements
B-8
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Command
Description
Page
time
Reports/sets current time (24-hour time)
B-38
up
Simulates pressing up pushbutton
B-26
zflow
Reports flow through zero air flow controller
B-8
avg time
This command reports the photometer averaging time in seconds. The
example below shows that the averaging time is 300 seconds, according to
Table B–2
. This command is only available if the photometer option is
installed.
Send:
avg
time
Receive:
avg time 5:300 sec
set avg time
selection
This command sets the averaging time according to
Table B–2
. The
example below sets the averaging time to 120 seconds. This command is
only available if the photometer option is installed.
Send:
set avg time 2
Receive:
set avg time 2 ok
Table B–2.
Averaging Times
Selection
Averaging Time (seconds)
0 60
seconds
1 90
2 120
3 180
4 240
5 300
gas conc
This command reports the current concentration of gas being supplied by
the Model146
i
.
Send:
gas
conc
Receive:
gas conc actual 100.000 target 1.600 ppm
gflow
zflow
These commands report the current zero air or gas flow.
Measurements
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...