Operation
Instrument Controls Menu
3-58
Model 146
i
Instruction Manual
Thermo Fisher Scientific
OZONATOR
PERM OVEN
USER 1
USER 2
USER 3
USER 4
USER 5
USER 6
USER 7
USER 8
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Setting
s
.
DIGITAL INPUT SETTINGS:
>1 NOP GAS BIT 1
2 NOP GAS BIT 2
3 NOP GAS BIT 3
4 NOP SPAN BIT 1
5 NOP SPAN BIT 2
6 NOP SPAN BIT 3
7 NOP O3/PERM BIT 1
ª
OPER MODE DIAGS ALARM
The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
and closes to trigger the digital input action. If nothing is connected to the
digital input pin, the state should be left at open to prevent the action from
being triggered.
●
Press
to toggle and set the logic state open or closed.
Digital Input Settings
Logic State
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...