Operation
Diagnostics Menu
3-76
Model 146
i
Instruction Manual
Thermo Fisher Scientific
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
OPER MODE DIAGS ALARM
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and normal operation.
The example below shows the selected output state “ALL” is set to normal.
●
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1-6
,
or
Current Channel 1-6
.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
OPER MODE DIAGS ALARM
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Note
If the analyzer is in service mode, pressing
on the item will
toggle it yes or no (with the exception of purchased options).
▲
●
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
Set Analog Outputs
Instrument
Configuration
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...