Operation
Photometer Setup Menu
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-33
outputs are updated every 5 seconds regardless of the averaging time. An
averaging time of 300 seconds means that the moving average
concentration of the last 300 seconds will be output every 5 seconds. The
lower the averaging time the faster the front panel display and analog
outputs respond to concentration changes. Longer averaging times are
typically used to smooth output data.
The following averaging times are available: 60, 90, 120, 180, 240, and
300 seconds.
●
In the Main Menu, choose Photometer Setup >
Averaging Time
.
AVERAGING TIME:
CURRENTLY: 90 SEC
SET TO: 120 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
OPER MODE DIAGS ALARM
The Calibration Factors menu displays the calibration factors and is used to
make minor compensations to the O
3
concentration.
●
In the Main Menu, choose Photometer Setup >
Calibration Factors
.
CALIBRATION FACTORS:
>O3 BKG 0.0000
O3 COEF 1.000
RESET USER CAL DEFAULTS
OPER MODE DIAGS ALARM
The O
3
Background screen is used to perform a manual adjustment of the
instruments zero background. Before performing a background adjustment,
allow the analyzer to sample zero air until stable readings are obtained.
●
In the Main Menu, choose Photometer Setup > Calibration Factors
>
O3 Bkg
.
Calibration Factors
O
3
Background
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...