System Description
Flow System
Thermo Fisher Scientific
Model 146
i
Instruction Manual
8-5
During standby operation, a continuous flow of gas is necessary through the
oven. The standby flow and operating flow should be the same to +10
percent. This is achieved by setting the pressure regulator to the same
pressure setting as the back pressure regulator. In practice, rather than
measuring the pressure, it is the actual flows that are measured while
adjusting the pressure regulator. Two solenoids are used to change the flow
of gas going through the oven to the main stream or to vent.
Photometer
The photometer has an airtight chamber that contains the sample and
reference gases and includes a photometer lamp at one end and a detector at
the other end.
Detector System
The photo tube in the detector transmits light intensity information to the
measurement interface board for sample measurement computations.
Pressure Transducer
The pressure transducer measures the pressure of the sample gas.
Ozonator
The internal ozonator operates on the photolytic principle and the ozone
level produced is a function of light intensity at 185 nm and gas flow. The
light intensity is varied by changing the current into the lamp. The gas flow
is held constant by a pressure regulator followed by a capillary.
Ozonator Lamp Power
Supply
The ozonator lamp power supply generates high voltage AC and contains
heater control circuits for the ozonator lamp.
Photometer Lamp Power
Supply
The photometer lamp power supply generates high voltage AC and contains
heater control circuits for the photometer lamp.
Photometer Temperature
Thermistor
The photometer temperature thermistor is used for controlling the
photometer at a consistent temperature.
Pump
The pump draws the sample gas through the photometer.
Sample/Reference
Solenoid Valves
The sample/reference solenoid valves allow sample gas or reference (zero)
gas to pass through the cell.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...