Operation
Alarms Menu
3-78
Model 146
i
Instruction Manual
Thermo Fisher Scientific
ALARMS:
ALARMS DETECTED 0
>AMBIENT TEMP OK
O3 LAMP TEMP OK
PERM OVEN TEMP OK
PRESSURE OK
PHOT BENCH TEMP OK
PHOT LAMP TEMP OK
ª
OPER MODE DIAGS ALARM
PHTO INTENSITY OK
MOTHERBOARD STATUS OK
INTERFACE STATUS OK
I/O EXP STATUS OK
The Ambient Temperature screen displays the current ambient temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 8 to 47 °C. If the ambient temperature reading goes beyond
either the minimum or maximum alarm limit, an alarm is activated. The
word “ALARM” appears in the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Ambient Temp
.
AMBIENT TEMPERATURE:
ACTUAL 30.1
o
C
>MIN 15.0
o
C
MAX 45.0
o
C
OPER MODE DIAGS ALARM
The Minimum Ambient Temperature alarm limit screen is used to change
the minimum ambient temperature alarm limit. The minimum and
maximum ambient temperature screens function the same way.
●
In the Main Menu, choose Alarms > Ambient Temp >
Min
or
Max
.
Ambient
Temperature
Min and Max Ambient
Temperature Limits
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...