Servicing
Fan Replacement - Rear Panel
Thermo Fisher Scientific
Model 146
i
Instruction Manual
7-17
5. Install the new pump by following the previous steps in reverse.
Fan Replacement -
Rear Panel
Use the following procedure to replace the rear panel fan (
).
Equipment Required:
Fan
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the measurement bench. Refer to
Bench and Lowering the Partition Panel”
.
3. Pull the power connector off the fan.
4. Remove the fan guard from the fan and remove the filter.
5. Remove the four fan mounting screws and remove the fan.
6. Install a new fan following the previous steps in reverse.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...