Preventive Maintenance
Pump Rebuilding
5-10
Model 146
i
Instruction Manual
Thermo Fisher Scientific
3. Remove the four screws from the top plate, remove top plate, flapper
valve, and the bottom plate.
Figure 5-5.
Rebuilding the Pump
4. Remove the screw the securing the diaphragm to piston and remove
diaphragm.
5. Assemble the pump by following the previous steps in reverse, make sure
the Teflon® (white) side of the diaphragm is facing up and that the
flapper valves cover the holes of the top and bottom plate.
6. Test the pump for leaks. Refer to “Leak Test and Pump Checkout” in
the “Servicing” chapter.
Top Plate
Flapper Plate
Bottom Plate
Diaphragm
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...