C-Link Protocol Commands
Hardware Configuration
B-38
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Level
Contrast Level
7 70%
8 80%
9 90%
10 100%
date
This command reports the current date. The example below reports the
date as December 1, 2004.
Send:
date
Receive: date
12-01-04
set date
mm-dd-yy
mm
= month
dd
= day
yy
= year
This command sets the date of the analyzer’s internal clock. The example
below sets the date to December 1, 2004.
Send:
set
date
12-01-04
Receive:
set date 12-01-04 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send:
set default params
Receive:
set default params ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example below saves the parameters to FLASH
memory.
Send:
set
save
params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example
below reports that the internal time is 2:15:30 pm.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...