C-Link Protocol Commands
Datalogging
B-16
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Send:
photometer
temp
Receive:
photometer temp 31.2 deg C, actual 31.2
program
This command reports programmed events status.
Send:
program
Receive: program
on
set program
onoff
This command sets programmed events
on
or
off
. If set to
off
, programmed
events will not be triggered at their normal interval and if programmed
events are currently running, they will be aborted. All the other set
commands will be disabled if the programmed events are running, except
for this command. If set to
on
, programmed events will be triggered at their
normal interval.
Send:
set program on
Receive:
set program on ok
clr records
This command will clear all long and short records that have been saved.
Send:
clear
records
Receive:
clear records ok
set clr lrecs
set clr srecs
These commands will clear only the long records or only the short records
that have been saved. The example below clears short records.
Send:
set
clr
srecs
Receive:
set clr srecs ok
list din
list dout
These commands report the current selection for the digital outputs in the
format. Output no Index number variable name active state. The active
state for digital outputs is open or closed. The active state for digital inputs
is high or low.
Send:
list
dout
Receive: list
dout
output
index
variable
state
1 1 GAS BIT 1 open
2 2 GAS BIT 2 open
3 3 GAS BIT 3 open
4 4 SPAN BIT 1 open
Datalogging
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...