Operation
Photometer Setup Menu
3-30
Model 146
i
Instruction Manual
Thermo Fisher Scientific
The Model 146
i
can be equipped with an optional internal photometer,
which is used to measure concentration of ozone.
The Photometer Setup menu sets up the photometer.
●
In the Main Menu, choose
Photometer Setup.
PHOTOMETER:
>RANGE
AVERAGING TIME
CALIBRATION FACTORS
CALIBRATION
OPER MODE DIAGS ALARM
The Range menu allows the operator to view the gas units (always in ppm),
to select the O
3
range, and to set the custom ranges.
●
In the Main Menu, choose Photometer Setup >
Range.
RANGE:
GAS UNITS PPM
>RANGE 5.00
SET CUSTOM RANGES
OPER MODE DIAGS ALARM
The O
3
Range screen defines the concentration range of the analog
outputs. For example, an O
3
range of 0–5 ppm sets the analog output to
concentrations between 0 and 5 ppm. Available operating ranges include:
0.05, 0.1, 0.2, 0.5, 1.0, 2.0, 5.0, C1, C2, and C3.
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “
Set Custom Ranges
” below.
The display shows the current O
3
range. The next line of the display is used
to change the range.
●
In the Main Menu, choose Range >
O3 Range
.
Photometer Setup
Menu
Range
O
3
Range
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...