Operation
Flow Modes
3-16
Model 146
i
Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Flow Modes >
O3.
O3 MODE SELECTION:
>OFF
TRANSFER STANDARD
GPT
PHOTOMETER CONTROL
OPER MODE DIAGS ALARM
FLOW MODE:
OZONATOR OFF
SAVING
SAVE FLOW MODE
OPER MODE DIAGS ALARM
Five ozonator levels (1-5) can be set using the Ozonator Level Selection
screen. Once selected the screen reports the new flow mode. The user must
press
to accept and activate these settings. The message “SAVING”
is displayed for 3 seconds, then removed.
●
In the Main Menu, choose Flow Modes > O3 > Transfer Standard >
Level 1
-
5.
O3 LEVEL SELECTION:
>LEVEL1 20.0 %
LEVEL2 25.0 %
LEVEL3 40.0 %
LEVEL4 60.0 %
LEVEL5 80.0 %
MANUAL 2.0 %
OPER MODE DIAGS ALARM
Transfer Standard
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...