Operation
Alarms Menu
3-80
Model 146
i
Instruction Manual
Thermo Fisher Scientific
The Permeation Oven Temperature displays the current permeation gas
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from25 to 50 °C. If the permeation oven temperature
reading goes beyond either the minimum or maximum limit, an alarm is
activated. The word “ALARM” appears in the Run screen and in the Main
Menu.
●
In the Main Menu, choose Alarms >
Perm Oven Temp
.
POVEN GAS TEMPERATURE:
ACTUAL 48.4
o
C
>MIN 40.0
o
C
MAX 50.0
o
C
OPER MODE DIAGS ALARM
The Minimum Permeation Oven Temperature alarm limit screen is used to
change the minimum bench temperature alarm limit. The minimum and
maximum permeation oven temperature screens function the same way.
●
In the Main Menu, choose Alarms > Perm Oven Temp >
Min
or
Max
.
POVEN GAS EMPERATURE:
ACTUAL MIN: 40.0
o
C
SET MIN TO: 44.0
o
C ?
©ª
INC/DEC
SAVE VALUE
OPER MODE DIAGS ALARM
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 200 to 1,000 mmHg. If the pressure reading goes beyond either
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Pressure
.
Permeation Oven
Temperature
Min and Max Permeation
Oven Temperature Limits
Pressure
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...