Operation
Service Menu
3-84
Model 146
i
Instruction Manual
Thermo Fisher Scientific
The Service menu appears only when the instrument is in the service mode.
To put the instrument into the service mode:
●
In the Main Menu, choose Instrument Controls >
Service Mode
.
Advanced diagnostic functions are included in the service mode.
Meaningful data should not be collected when the instrument is in the
service mode.
●
In the Main Menu, choose
Service
.
SERVICE:
>FLOW CONTROL SELECTION
CALIBRATION PRES/TEMP
ZERO FLOW CALIBRATION
GAS FLOW CALIBRATION
PRESSURE CALIBRATION
AMBIENT TEMP CALIBRATION
BENCH TEMP CALIBRATION
ª
OPER MODE DIAGS ALARM
ANALOG OUT CALIBRATION
ANALOG INPUT CALIBRATION
PERM OVEN SETTINGS
LAMP SETTING
INTENSITY CHECK
O3 RESPONSE COEFFICIENT
DISPLAY PIXEL TEST
RESTORE USER DEFAULTS
The Flow Control Selection menu is used to select the installed zero-air
controller and gas controller. The zero-air and gas controller screens are
used when installing new mass flow controllers (for example, when
installing controllers other than the standard 10 slm for zero-air, or 100
sccm for gas). The flow control selection is visible only when the
instrument is in service mode. For more information on the service mode,
see “
Service Mode
” earlier in this chapter.
●
In the Main Menu, choose Service >
Flow Control Selection.
Service Menu
Flow Control Selection
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...