C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-13
example below sets the ozonator lamp temperature alarm maximum value
to 58 °C. These commands are only available if the photometer option is
installed.
Send:
set alarm o3 lamp temp max 58.0
Receive:
set alarm o3 lamp temp max 58.0 ok
alarm pressure min
alarm pressure max
These commands report the photometer bench pressure alarm minimum
and maximum value current settings. The example below reports that the
pressure alarm minimum value is 200 mmHg. These commands are only
available if the photometer option is installed.
Send:
alarm pressure min
Receive:
alarm pressure min 200 mm_Hg
set alarm pressure min
value
set alarm pressure max
value
These commands set the photometer bench pressure alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing pressure alarm limits in millimeters of mercury. The example
below sets the pressure alarm maximum value to 1000 mmHg. These
commands are only available if the photometer option is installed.
Send:
set alarm pressure max 1000
Receive:
set alarm pressure max 1000 ok
alarm trig conc o3
This command reports the photometer O
3
concentration alarm trigger
action for minimum alarm to either floor or ceiling. The example below
shows the O
3
concentration minimum alarm trigger to ceiling, according to
Table B–3
. These commands are only available if the photometer option is
installed.
Send:
alarm trig conc o3
Receive:
alarm trig conc o3 1
set alarm trig conc o3
value
These commands set the photometer O
3
concentration alarm trigger action
for the minimum alarm to
value
, where
value
is set to either floor or
ceiling, according to
Table B–3
. The example below sets the O
3
concentration minimum alarm trigger to floor. These commands are only
available if the photometer option is installed.
Send:
set alarm trig conc 0
Receive:
set alarm trig conc 0 ok
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...