C-Link Protocol Commands
Measurement Configuration
B-36
Model 146
i
Instruction Manual
Thermo Fisher Scientific
custom
range
range
= | 1 | 2 | 3 |
This command reports the user-defined value of custom
range
1, 2, or 3.
The example below reports that custom range 1 is defined to 4.50 ppm.
This command is only available if the photometer option is installed.
Send:
custom
1
Receive:
custom 1 4.500E+03 ppb
set custom
range
range
value
set custom 1 range
value
set custom 2 range
value
set custom 3 range
value
These commands are used to set the maximum concentration for any of the
three custom
ranges
1, 2, or 3 to range
value
, where
value
is a floating-point
number representing concentration in ppm. The example below sets the
custom 1 range to 4.5 ppm.
Send:
set custom 1 range 4.5
Receive:
set custom 1 range 4.5 ok
pres comp
This command reports whether pressure compensation is on or off. The
example below shows that pressure compensation is on. This command is
only available if the photometer option is installed.
Send:
pres
comp
Receive: pres
comp
on
set pres comp
onoff
These commands turn the pressure compensation
on
or
off
. The example
below turns pressure compensation off. This command is only available if
the photometer option is installed.
Send:
set pres comp off
Receive:
set pres comp off ok
span index
This command reports the current span index, where 0 is zero air, 1-5 are
pre-set span levels, and 6 is manual span level.
Send:
span
index
Receive:
span index 1
set span index
index
index
= 0-6
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...