Operation
Flow Modes
Thermo Fisher Scientific
Model 146
i
Instruction Manual
3-15
The Manual Gas Zero Air Flow screen sets the manual zero-air flow value
for the previously selected gas. The screen identifies the minimum and
maximum acceptable zero-air flow rates and lets the user select a flow rate
between these two settings. If the selected zero-air flow is lower or higher
than the acceptable range, “TOO LOW” or TOO HIGH” appears on the
screen.
●
In the Main Menu, choose Flow Modes > Permeation Oven > Gas Mix
> Gas A-F > Manual >
Zero Air.
MAN GAS A ZERO AIR FLOW:
MINIMUM: 00400 SCCM
MAXIMUM: 10000 SCCM
SET TO: 0500
1
SCCM?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Manual Gas Flow screen sets the manual gas flow value for the
previously selected gas. The screen identifies the minimum and maximum
acceptable gas flow rates and lets the user select a flow rate between these
two settings. If the selected gas flow is lower or higher than the acceptable
range, “TOO LOW” or TOO HIGH” appears on the screen.
●
In the Main Menu, choose Flow Modes > Permeation Oven > Gas Mix
> Gas A-F > Manual >
Gas.
MAN GAS A GAS FLOW:
MINIMUM: 004.00 SCCM
MAXIMUM: 100.00 SCCM
SET TO: 050.0
1
SCCM?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The O
3
flow mode menu is used to set the operational mode of the
ozonator (if the ozonator option is installed). Selecting OFF reports the
new flow mode. The user must press
to accept and activate these
settings. The message “SAVING” is displayed for 3 seconds, then removed.
Zero Air
Gas
O
3
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...