Servicing
Reaction Chamber/Mixing Chamber Replacement
7-48
Model 146
i
Instruction Manual
Thermo Fisher Scientific
Figure 7-15.
Replacing the Glass Chamber Assembly
Reaction
Chamber/Mixing
Chamber
Replacement
Use the following procedure to replace the reaction chamber or mixing
chamber.
Equipment Required:
Reaction chamber or mixing chamber
Allen wrench, 7/64-inch
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the plumbing.
3. Remove two (2) Allen screws from the top of the holder assembly.
4. Remove two (2) Allen screws from the side of the holder assembly.
Permeation
Oven Top
See Expanded View
Expanded View
Chamber Top
Oven Top
Glass
Chamber
Assembly
Permeation
Oven Bottom
See Procedure
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...