C-Link Protocol Commands
Commands
B-2
Model 146
i
Instruction Manual
Thermo Fisher Scientific
The instrument must be in the remote mode in order to change instrument
parameters via remote. However, the command “set mode remote” can be
sent to the analyzer to put it in the remote mode. Report commands
(commands that don’t begin with “set”) can be issued either in the remote
or local mode. For information on changing modes, see Chapter 3,
“
Operation
”.
The commands can be sent in either uppercase or lowercase characters.
Each command must begin with the proper instrument identification
number (ASCII) character. The command in the example below begins
with the ASCII character code 179 decimal, which directs the command to
the Model 146
i
, and is terminated by a carriage return “CR” (ASCII
character code 13 decimal).
<ASCII 179> T I M E <CR>
If an incorrect command is sent, a “bad command” message will be
received. The example below sends the incorrect command “set unit ppm”
instead of the correct command “set gas unit ppm.”
Send:
set
unit
ppm
Receive:
set unit ppm bad cmd
The “save” and “set save params” commands stores parameters in FLASH.
It is important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure.
Table B–1
lists the 146
i
C-Link protocol commands. The interface will
respond to the command strings outlined below.
Table B–1.
C-Link Protocol Commands
Command
Description
Page
addr dns
Reports/sets domain name server address for Ethernet port
B-39
addr gw
Reports/sets default gateway address for Ethernet port
B-39
addr ip
Reports/sets IP address for Ethernet port
B-39
addr nm
Reports/sets IP netmask address for Ethernet port
B-40
alarm bench lamp
temp max
Reports/sets photometer bench lamp temperature alarm
maximum value
B-10
Commands
Commands List
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...