Operation
Service Menu
3-88
Model 146
i
Instruction Manual
Thermo Fisher Scientific
ZERO AIR DRIVE 5%:
STD FLOW : 00.500 SLM
VOL FLOW: 00.50
1
LPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
OPER MODE DIAGS ALARM
The Gas Flow Calibration menu is used to calibrate the gas flow controller
at 5, 20, 35, 50, 65, 80, and 95 percent of full-scale. Using a NIST
traceable flow meter, the accuracy of the controller can be calibrated to
about
±
1%, as opposed to the
±
2% accuracy rating when shipped from
the manufacturer. The gas flow calibration is visible only when the
instrument is in service mode. For more information of the service mode,
see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Gas Flow Calibration.
GAS FLOW CAL:
>SOLENOID A
5% FS 0.500 SCCM
20% FS 2.000 SCCM
35% FS 3.500 SCCM
50% FS 5.000 SCCM
65% FS 6.500 SCCM
80% FS 8.000 SCCM
OPER MODE DIAGS ALARM
95% FS 9.500 SCCM
The Gas Flow Calibration Solenoid Select is used to select the solenoid that
is active during the gas flow controller calibration. Solenoid choices
include: Solenoid A, B, or C, and Solenoid D, E, or F (if installed).
●
In the Main Menu, choose Service > Gas Flow Calibration >
Solenoid.
Gas Flow Calibration
Gas Flow Calibration
Solenoid Select
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...