Calibration
Ozonator Control Check
Thermo Fisher Scientific
Model 146
i
Instruction Manual
4-3
Figure 4-1.
Photometer Calibration Flow Schematic
Ozonator Control
Check
Use the following procedure to perform an ozonator control check. Refer to
the ozonator control check flow schematic (
1. Connect the 146
i
OUTPUT port to the 49
i
PS INPUT port.
2. Set ozone concentration at different levels.
3. Compare the ozone concentration readings on the 146
i
and 49
i
PS.