C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-45
This command assigns digital input
channel
(1-16) to activate the action
indicated by
index
(1-35), when the input transitions to the designated
state
(high or low). Use “list din var” command to obtain the list of supported
index
values and corresponding actions.
Send:
set din 5 9 high
Receive:
set din 1 9 high ok
dout
channel
This command reports the index number and output variable and the
active state assigned to output
channel
. The example below reports the
input 2 to be assigned an index number 2 corresponding to “local/remote”
with the active state being open.
Send:
dout
2
Receive:
dout 2 2 LOCAL/REMOTE open
set dout
channel index state
This command assigns digital output
channel
to be assigned to the action
associated with
index
, and assigns it an active state of
state
(open or closed).
Send:
set dout 2 2 open
Receive:
set dout 2 2 open ok
dtoa
channel
This reports the outputs of the 6 or 12 Digital to Analog converters,
according to
Table B–13
. The example below shows that the D/A #1 is
97.7% full-scale.
Send:
dtoa
1
Receive: dtoa
1
97.7%
Note
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
▲
Table B–13.
Default Output Assignment
D to A
Function
Single Range
Dual Range
Auto Range
1 Voltage
Output
O
3
Not Used
Not Used
2 Voltage
Output
O
3
Not Used
Not Used
3
Voltage Output
Not Used
Not Used
Not Used
4
Voltage Output
Not Used
Not Used
Not Used
5
Voltage Output
Not Used
Not Used
Not Used
6
Voltage Output
Not Used
Not Used
Not Used
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...