7. Replace the measurement bench.
Figure 7-6.
Replacing the Rear Panel Fan
Fan Replacement -
Mass Flow
Controllers
Use the following procedure to replace the cooling fan for the mass flow
controllers (
Equipment Required:
Fan
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the mass flow controllers. Refer to
3. Disconnect the electrical connection from the fan.
4. Remove the fan mounting screws (4) and nuts and remove the fan.
5. Install the new fan using the previous steps in reverse.
Fan Guard
Filter
Mounting Screws (4)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...