Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-188
2470-901-01 Rev. A /
May
2019
Trigger events
Event description
Event constant
Front-panel TRIGGER key press
DISPlay
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender
<n>
(up to two), which combines trigger
events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Source limit condition occurs
SLIMit
Example
*RST
SENS:FUNC "CURR"
TRIG:LOAD "LoopUntilEvent", DISP, 25
INIT
Reset the instrument.
Set the instrument to measure DC current.
Set the LoopUntilEvent trigger model to make
measurements until the front-panel TRIGGER key is
pressed after starting the trigger model, then make
measurements that constitute 75% of the reading buffer.
Start the trigger model.
Also see
:TRIGger:LOAD "SimpleLoop"
This command loads a predefined trigger model configuration.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD "SimpleLoop", <count>
:TRIGger:LOAD "SimpleLoop", <count>, <delay>
:TRIGger:LOAD "SimpleLoop", <count>, <delay>, "<bufferName>"
<count>
The number of measurements the instrument will make
<delay>
The delay time before each measurement (167 ns to 10 ks); default is
0
for no delay
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used