:TRIGger:TIMer<n>:STARt:STIMulus
This command describes the event that starts the trigger timer.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NONE
Usage
:TRIGger:TIMer<n>:STARt:STIMulus <event>
:TRIGger:TIMer<n>:STARt:STIMulus?
<n>
Trigger timer number (1 to 4)
<event>
The event that starts the trigger timer; see
Details
Details
Set this command to any trigger event to start the timer when that event occurs.
Set this command to none to disable event processing and use the timer as a timer or alarm based
on the start time.
Trigger events are described in the table below.
Trigger events
Event description
Event constant
Trigger event blender
<n>
(up to two), which combines trigger
events
BLENder<n>
A command interface trigger:
Any remote interface: *
TRG
GPIB only: GET bus command
USB only: A USBTMC TRIGGER message
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
Front-panel TRIGGER key press
DISPlay
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
No trigger event
NONE
Notify trigger block
<n>
(1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
Source limit condition occurs
SLIMit
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Line edge detected on TSP-Link synchronization line
<n>
(1 to 3)
TSPLink<n>