Section 3: Instrument description
2470 High Voltage SourceMeter Instrument Reference Manual
3-46
2470-901-01 Rev. A /
May
2019
Using TSP commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
createconfigscript("
setupName
")
Where
setupName
is the name of the setup script that is created.
Save a user setup to a USB flash drive
You cannot save to the flash drive using SCPI commands.
From the front panel:
1.
Save the user setup to internal memory, as described in
Save a user setup to internal memory
(on page 3-45).
2.
Insert the USB flash drive into the USB port on the front panel.
3.
Press the
MENU
key.
4.
Under Scripts, select
Manage
. The MANAGE SCRIPTS window is displayed.
5.
In the Internal Scripts list, select the script you want to copy to the USB flash drive.
6.
Select
>
. The file is transferred to the USB flash drive, and the corresponding filename is
displayed in the USB Scripts box.
Using TSP commands:
1.
Save the user setup to internal memory, as described in
Save a user setup to internal memory
(on page 3-45).
2.
Insert the USB flash drive into the USB port on the front panel.
3.
Send the command:
setupName
.save("/usb1/
USBSetupName
")
Where
setupName
is the name of the user setup and
USBSetupName
is the name of the file on
the USB flash drive. You can use the same name for
setupName
and
USBSetupName
.
Copy a user setup
To copy a user setup from an external USB flash drive to the instrument from the front panel:
1.
Insert the USB flash drive into the USB port on the front panel.
2.
Press the
MENU
key.
3.
Under Scripts, select
Manage
. The MANAGE SCRIPTS window is displayed.
4.
In the USB Scripts list, select the script you want to copy from the USB flash drive.
5.
Select
<
. The file is transferred to the instrument, and the corresponding filename is displayed in
the Internal Scripts box.