Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-174
2470-901-01 Rev. A /
May
2019
:TRIGger:DIGital<n>:OUT:PULSewidth
This command describes the length of time that the trigger line is asserted for output triggers.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10e-6 (10 µs)
Usage
:TRIGger:DIGital<n>:OUT:PULSewidth <width>
:TRIGger:DIGital<n>:OUT:PULSewidth?
<n>
Digital I/O trigger line (1 to 6)
<width>
Pulse length (0 to 100 ks)
Details
Setting the pulse width to zero (0) seconds asserts the trigger indefinitely.
Example
DIG:LINE1:MODE TRIG, OUT
TRIG:DIG1:OUT:PULS 2
Set digital line 1 to trigger out.
Set the pulse to 2 s.
Also see
Digital I/O port configuration
:TRIGger:DIGital<n>:OUT:STIMulus
This command selects the event that causes a trigger to be asserted on the digital output line.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
NONE
Usage
:TRIGger:DIGital<n>:OUT:STIMulus <event>
:TRIGger:DIGital<n>:OUT:STIMulus?
<n>
Digital I/O trigger line (1 to 6)
<event>
The event to use as a stimulus; see
Details