2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-275
Details
The binary equivalent of the returned value indicates the input pattern on the I/O port. The least
significant bit of the binary number corresponds to line 1 and the value of bit 3 corresponds to line 3.
For example, a returned value of 2 has a binary equivalent of 010. This indicates that line 2 is high
(1), and that the other two lines are low (0).
Example
data = tsplink.readport()
print(data)
Reads state of all three TSP-Link lines.
Assuming line 2 is set high, the output is:
2.00
(binary 010)
The format of the output may vary depending on the ASCII
precision setting.
Also see
Triggering using TSP-Link trigger lines
(on page 9-6)
(on page 14-273)
tsplink.state
This attribute describes the TSP-Link online state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
state
= tsplink.state
state
TSP-Link state (
online
or
offline
)
Details
When the instrument power is first turned on, the state is
offline
. After
tsplink.initialize()
or
tsplink.reset()
is successful, the state is
online
.
Example
state = tsplink.state
print(state)
Read the state of the TSP-Link system. If it is online, the output
is:
online
Also see
(on page 14-270)
(on page 14-274)