2470 High Voltage SourceMeter Instrument Reference Manual
Section 4: Sourcing and measuring
2470-901-01 Rev. A /
May
2019
4-103
Using TSP commands:
To delete configuration index 8 from a source configuration list named
MyConfigList
, send the
command:
smu.source.configlist.delete("MyConfigList", 8)
For a measure configuration list, replace
source
with
measure
.
To delete an entire source configuration list named
MyConfigList
, send the command:
smu.source.configlist.delete("MyConfigList")
For a measure configuration list, replace
source
with
measure
.
Viewing the available configuration lists
You can use remote commands to view the names of the configuration lists stored on the instrument.
Using SCPI commands:
To receive the name of one source configuration list stored on the instrument, use the following
command.
:SOURce:CONFiguration:LIST:CATalog?
For a measure configuration list, replace
:SOURce
with
:SENSe
.
Each time this command executes, the name of one defined configuration is returned. Keep sending
this command until it returns an empty string to get all defined lists. After the command returns an
empty string, it wraps around and starts returning names again. If only an empty string is returned, no
configuration lists of the specified type exist.
Using TSP commands:
To receive the name of one source configuration list stored on the instrument, use the following
command.
print(smu.source.configlist.catalog())
For a measure configuration list, replace
source
with
measure
.
Each time this command executes, the name of one defined configuration is returned. Keep sending
this command until it returns
nil
to get all defined lists. After the command returns
nil
, it wraps
around and starts returning names again. If only
nil
is returned, no configuration lists of the specified
type exist.