Section 4: Sourcing and measuring
2470 High Voltage SourceMeter Instrument Reference Manual
4-60
2470-901-01 Rev. A /
May
2019
To create a sweep:
1.
Set the source function using
:SOURce[1]:FUNCtion[:MODE]
.
2.
Set the source range using
:SOURce[1]:<function>:RANGe
.
3.
Set any other source settings that apply to your sweep. You must set source settings before the
sweep function is called.
4.
If you are using
:SOURce[1]:SWEep:<function>:LIST
, set up the source configuration list
for your sweep.
5.
Set the parameters for the sweep command.
6.
Set the measurement function using
[:SENSe[1]]:FUNCtion
.
7.
Set the measurement range using
[:SENSe[1]]:<function>:RANGe[:UPPer]
.
8.
Make any other settings appropriate to your sweep.
9.
Send
:INITiate
to start the sweep.
To save your settings, save them to a user-saved setup using the *
SAV
command.
For example sweeps, see
(on page 4-63).
For detail on the commands and options listed above, see the following command descriptions:
•
(on page 12-67)
•
[:SENSe[1]]:<function>:RANGe[:UPPer]
(on page 12-53)
•
(on page 12-80)
•
(on page 12-82)
•
:SOURce[1]:CONFiguration:LIST:CREate
(on page 12-68)
•
:SOURce[1]:CONFiguration:LIST:STORe
(on page 12-72)
•
:SOURce[1]:SWEep:<function>:LINear
(on page 12-88)
•
:SOURce[1]:SWEep:<function>:LINear:STEP
(on page 12-90)
•
:SOURce[1]:SWEep:<function>:LIST
(on page 12-92)
•
:SOURce[1]:SWEep:<function>:LOG
(on page 12-94)