2470 High Voltage SourceMeter Instrument Reference Manual
Section 3: Instrument description
2470-901-01 Rev. A /
May
2019
3-47
Delete a user setup
To remove a user setup from internal memory or the USB flash drive from the front panel:
1.
Press the
MENU
key.
2.
Under Scripts, select
Manage
. The MANAGE SCRIPTS window is displayed.
3.
Under Internal Scripts or USB Scripts, select the name of the script.
4.
Select
Delete
. A confirmation message is displayed.
5.
Select
OK
.
To delete a user setup from internal memory using SCPI commands:
You must overwrite an existing setup with the new setup. See
Save a user setup to internal memory
(on page 3-45).
To delete a user setup from internal memory using TSP commands:
To delete the setup, send the command:
script.delete("
setupName
")
Where
setupName
is the name of the script that will be deleted.
Recall a user setup
You can recall setups from internal nonvolatile memory or a USB flash drive. When you recall a setup,
you run a script that restores the instrument to the settings that are saved in that script.
To recall a saved setup from the front panel:
1.
Press the
MENU
key.
2.
Under Scripts, select
Run
.
3.
In the Available Scripts list, select the script you to recall. USB scripts have the prefix
usb1/
.
4.
Select
Run Selected
.
To recall a user setup from internal memory using SCPI commands:
Send the command:
*RCL <n>
Where
<n>
is an integer from 0 to 4 that represents the saved script.
To recall a saved setup using TSP commands:
Send the command:
setupName
()
Where
setupName
is the name of the script that contains the setup that was saved with
createconfigscript()
.