2470 High Voltage SourceMeter Instrument Reference Manual
Section 3: Instrument description
2470-901-01 Rev. A /
May
2019
3-25
Limit
Setting
Description
Limit 1 and Limit 2
These settings enable or disable limit testing. The Limit options
allow you to do pass-or-fail limit testing using the front panel of the
instrument. When you do a limit test, the home screen displays the
pass or fail result of the test.
Limit State enables or disables a limit test for the selected
measurement function. When testing is enabled, limit testing
occurs on each measurement. Limit testing compares the
measurements to the high and low limit values. If a measurement
is outside these limits, the test fails. If a measurement is in the
limits, it passes.
Config
Displays the settings that are available for the limit functions.
Audible
This determines if the instrument beeper sounds when a limit test
passes or fails.
Auto Clear
Auto clear indicates if the high and low limits should be cleared
automatically or not.
High Value
The Limit High Level specifies the upper limit for a limit test.
Low Value
The Low Value specifies the lower limit for limit tests.
Measure Config Lists menu
The
Config Lists
menu allows you to select an existing measure configuration list,
create a new list, load configuration settings to and from the instrument (system), and
view the settings of an index in a configuration list. For more information about using
configuration lists, see
(on page 4-83).
Setting
Description
Delete Index
Deletes a configuration list index from the selected configuration
list.
Index Details
Displays the details of the selected configuration index. Details
include settings such as function, value, delay, calculations, and
range. You can scroll the displayed list to view additional settings.
Index to System
Restores the instrument to the settings stored in the selected
configuration list index.
New List
Creates a new, empty configuration list. To populate the list with
the present instrument settings, select
System to Index
.
Select List
Displays a list of available configuration lists from which you can
choose. If you selected
Source
on the main menu, a menu of the
source configuration lists presently available on the instrument is
displayed.
If you selected
Measure
on the main menu, a menu of the
measure configuration lists presently available on the instrument is
displayed.
System to Index
Saves the present instrument settings to an index at the end of the
selected configuration list.