2470 High Voltage SourceMeter Instrument Reference Manual
Section 4: Sourcing and measuring
2470-901-01 Rev. A /
May
2019
4-43
Setting autozero
You can enable or disable automatic referencing. You can also request a one-time refresh of the
reference values.
The reference setting is stored with the selected measure function.
To set autozero using the front panel:
1.
Press the
FUNCTION
key.
2.
Select the source and measure combination.
3.
Press the
MENU
key.
4.
Under Measure, select
Settings
.
5.
For Auto Zero, select
On
or
Off
.
6.
If Off is selected, you can select the
Once
option to send a one-time refresh.
7.
Select
HOME
to return to the operating display.
To set autozero using SCPI commands:
Refer to the following commands:
•
[:SENSe[1]]:<function>:AZERo[:STATe]
(on page 12-46)
•
(on page 12-59)
To set autozero using TSP commands:
Refer to the following commands:
•
(on page 14-113)
•
(on page 14-114)