2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-59
[:SENSe[1]]:<function>:UNIT
This command sets the units of measurement that are displayed on the front panel of the instrument and stored in
the reading buffer.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Current: AMP
Voltage: VOLT
Usage
[:SENSe[1]]:<function>:UNIT <unitOfMeasure>
[:SENSe[1]]:<function>:UNIT?
<function>
The measure function:
Current:
CURRent[:DC]
Voltage:
VOLTage[:DC]
<unitOfMeasure>
Current:
OHM
,
WATT
, or
AMP
Voltage:
OHM
,
WATT
, or
VOLT
Details
The change in measurement units is displayed when the next measurement is made. You can only
change the units for the listed functions.
Example
VOLT:UNIT WATT
Changes the front-panel display and buffer readings for voltage
measurements to be displayed as power readings in watts.
Also see
None
[:SENSe[1]]:AZERo:ONCE
This command causes the instrument to refresh the reference and zero measurements once.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:AZERo:ONCE
Details
This command forces a refresh of the reference and zero measurements that are used for the present
aperture setting for the selected function.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the
drift, you can send the once command to make a reference and zero measurement immediately
before a test sequence.