Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-134
2470-901-01 Rev. A /
May
2019
Example
See the example in
(on page 14-131).
Also see
(on page 14-130)
smu.measure.math.enable
This attribute enables or disables math operations on measurements for the selected measurement function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
value
= smu.measure.math.enable
smu.measure.math.enable =
value
value
The math enable setting:
Disable:
smu.OFF
Enable:
smu.ON
Details
When this command is set to on, the math operation specified by the math format command is
performed before completing a measurement.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format = smu.MATH_PERCENT
smu.measure.math.enable = smu.ON
When voltage measurements are made, the math
format is enabled and set to percent.
Also see
Calculations that you can apply to measurements
(on page 4-49)
(on page 14-135)