Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-6
2470-901-01 Rev. A /
May
2019
The
STATus
buffer element returns status values for the readings in the buffer. The status values are
integers that encode the status value. Refer to the following table for values.
Buffer status bits for sense measurements
Bit (hex) Name
Decimal Description
0x0001 STAT_QUESTIONABLE
1
Measure status questionable
0x0006 STAT_ORIGIN
6
A/D converter from which reading originated; for the
2470, this will always be 0 (main)
0x0008 STAT_TERMINAL
8
Measure terminal, front is 1, rear is 0
0x0010 STAT_LIMIT2_LOW
16
Measure status limit 2 low
0x0020 STAT_LIMIT2_HIGH
32
Measure status limit 2 high
0x0040 STAT_LIMIT1_LOW
64
Measure status limit 1 low
0x0080 STAT_LIMIT1_HIGH
128
Measure status limit 1 high
0x0100 STAT_START_GROUP
256
First reading in a group
Example
TRACe:MAKE "voltMeasBuffer", 10000
MEAS:VOLT? "voltMeasBuffer", FORM, DATE, READ
Create a buffer named
voltMeasBuffer
. Make a voltage measurement and store it in the buffer
voltMeasBuffer
and return the formatted reading, the date, and the reading elements from the buffer.
Example output:
-00.0024 mV,05/16/2018,-2.384862E-06
Also see
(on page 12-6)
(on page 12-67)
(on page 12-120)
:READ?
This query makes measurements, places them in a reading buffer, and returns the last reading.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:READ?
:READ? "<bufferName>"
:READ? "<bufferName>", <bufferElements>
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1
is used
<bufferElements>
See
Details
; if nothing is specified,
READing
is used
Details
This query makes the number of readings specified by
[:SENSe[1]]:COUNt
. If multiple readings
are made, all readings are available in the reading buffer. However, only the last reading is returned
as a reading with the command. To get multiple readings, use the
:TRACe:DATA?
command.