2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-145
smu.measure.rel.enable
This attribute enables or disables the application of a relative offset value to the measurement.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
relEnable
= smu.measure.rel.enable
smu.measure.rel.enable =
relEnable
relEnable
Relative measurement control:
Disable relative offset:
smu.OFF
Enable relative offset:
smu.ON
Details
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset value calculated when you acquire the relative offset value.
Each returned measured relative reading is the result of the following calculation:
Displayed reading = Actual measured reading - Relative offset value
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
rel_value = smu.measure.rel.acquire()
smu.measure.rel.enable = smu.ON
Acquires a relative offset level value for voltage
measurements and turns the relative offset
feature on.