2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-133
smu.measure.limit[Y].high.value
This attribute specifies the upper limit for a limit test.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
1
Usage
highLimit
= smu.measure.limit[
Y
].high.value
smu.measure.limit[
Y
].high.value =
highLimit
highLimit
The value of the high limit (
−
9.11 to +9.11)
Y
Limit number:
1
or
2
Details
This command sets the high limit for the limit
Y
test for the selected measurement function. When limit
Y
testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Example
See the example in
(on page 14-131).
Also see
(on page 14-130)
smu.measure.limit[Y].low.value
This attribute specifies the lower limit for limit tests.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
−
1
Usage
lowLimit
= smu.measure.limit[
Y
].low.value
smu.measure.limit[
Y
].low.value =
lowLimit
lowLimit
The low limit value of limit
Y
(-9.11 to 9.11)
Y
Limit number:
1
or
2
Details
This command sets the lower limit for the limit
Y
test for the selected measure function. When limit
Y
testing is enabled, this causes a fail indication to occur when the measurement value is less than this
value.