2470 High Voltage SourceMeter Instrument Reference Manual
Section 4: Sourcing and measuring
2470-901-01 Rev. A /
May
2019
4-97
Storing configuration index 1
Use one of the following methods to:
•
Set the instrument source function to voltage
•
Set the instrument source range to 20 V
•
Set the instrument source limit to 2.0 V
•
Store all active source settings to
MySourceList
as configuration index 1 by appending to the
end of the initially empty list
Using SCPI commands:
:SOURce:FUNC VOLTage
:SOURce:VOLtage:RANGe 20
:SOURce:VOLTage:LEVel 2
:SOURce:CONF:LIST:STORe "MySourceList"
Using TSP commands:
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.source.level = 2
smu.source.configlist.store("MySourceList")
The following figure shows the active state of the instrument after you change the source level to
Instrument settings stored in a source configuration list
(on page 4-88) for a complete
list of source settings that the instrument stores in a source configuration list.
Figure 83: Example configuration point 1