2470 High Voltage SourceMeter Instrument Reference Manual
Section 8: Triggering
2470-901-01 Rev. A /
May
2019
8-33
Front panel options
When you select the Notify block from the front panel, the following options are available.
Setting
Description
Notify
The line or timer that is notified:
Digital Out Line: Select the digital output event (line 1
through line 6)
TSP-Link Out Line: Select the TSP-Link output event (line 1
through 3)
Timer: Select the timer event (timer 1 through 4)
LAN Out: Select the LAN output line (line 1 to 8)
Config
The available settings depend on the Notify selection. For the
digital, TSP-Link, and LAN out lines, select the pulse logic
(negative or positive).
Log event block
This block allows you to log an event in the event log when trigger model execution reaches this block.
You can also force the trigger model to abort with this block. When the trigger model executes the
block, the defined event is logged. If the abort option is selected, the trigger model is also aborted
immediately.
You can define the type of event (information, warning, abort model, or error). All events generated by
this block are logged in the event log. Warning and error events are also displayed in a popup on the
front-panel display.
Note that using this block too often in a trigger model could overflow the event log. It may also take
away from the time needed to process more critical trigger model blocks.
When you select the Log Event block, the following options are available.
Setting
Description
Event Type
The event number or type:
Abort Model: Stop the trigger model and log a warning
message
Information
N
: Logs an information message in the event log
Warning
N
: Logs a warning in the event log
Error
N
: Logs an error in the event log
Where
N
is 1 to 4; you can define up to four of each type.
Message
A message that you define.