Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-256
2470-901-01 Rev. A /
May
2019
trigger.timer[N].reset()
This function resets trigger timer settings to their default values.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.timer[
N
].reset()
N
Trigger timer number (1 to 4)
Details
The
trigger.timer[
N
].reset()
function resets the following attributes to their default values:
trigger.timer[
N
].count
trigger.timer[
N
].delay
trigger.timer[
N
].delaylist
trigger.timer[
N
].enable
trigger.timer[
N
].start.fractionalseconds
trigger.timer[
N
].start.generate
trigger.timer[
N
].start.seconds
trigger.timer[
N
].stimulus
It also clears
trigger.timer[
N
].overrun
.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 to their default
values.
Also see
(on page 14-252)
(on page 14-254)
(on page 14-255)
trigger.timer[N].start.fractionalseconds
(on page 14-257)
trigger.timer[N].start.generate
(on page 14-257)
trigger.timer[N].start.overrun
(on page 14-258)
trigger.timer[N].start.seconds
(on page 14-259)
trigger.timer[N].start.stimulus
(on page 14-259)